HETROC VACUUM GAUGE

HBS-J1 built with Pirani professional vacuum sensor can precisely measure the vacuum level. From the measurement, you can therefore gather information about humidity and the removal of foreign matter (oil, foreign gases, etc.), and thus know the best status to fill refrigerants. VVM-I can be used for monitoring vacuum system, leaking detection and distinguish the pump in quality.

Category: Brand:

Description

Safety Instruction

  • 2.1 Operation should according to this instruction and within the scope of the specified parameters.
  • 2.2 Replace the batteries with new ones when the power is low. Do not mix new batteries with old batteries, and different brands of batteries are forbidden. Remove all batteries if the instrument remains unused for a long period.
  • 2.3 Do not clean the instrument with corrosive detergents or solvents.
  • 2.4 Wear goggles and protective gloves.
  • 2.5 Tighten copper cap and stored in a dry place.

Operations

  • 5.1 Install three “AA” batteries.
  • 5.2 Long press and hold the power button ⏻ for 3 seconds to turn LCD light on. The display shows “—-” when the unit is completely warm-up.
  • 5.3 Connected the system directly or via connection accessories.
  • 5.4 Start the vacuum pump, the display will show ultimate pressure from high to low. Once reached the target ultimate pressure, the vacuum meter buzzer alarms and the backlight will blink simultaneously for 10 seconds.
  • 5.5 Press and hold power button ⏻ for 3 seconds to turn the power off.

 

ParameterSpecification
Maximum Overload Pressure14PSI/0.1Mpa
Range0–10000 Pa, 0–100.00 mBar, 0–75.000 mmHg, 0–75000 Micron
Resolution0.01 (less than 10 Pa), 0.0001 (less than 10 mBar), 0.0001 (less than 10 mmHg), 1 (less than 30000 Microns)
Accuracy2–100 Pa: ±5% of reading (at 20°C)
Operating Temperature0–50°C (32–122°F)
Battery Life45 hours (3xAA batteries)
UnitPa, mBar, mmHg, Microns
Refresh Rate0.5 second
Connections1×1/4” Flare
Sensor1x Pirani sensor
Auto Off Time10 minutes
Backlight Time20 seconds
Buzzer Alarm90 dB, shut down after 10 seconds
ApplicationAtmospheric environment (It cannot be measured in the presence of refrigerants and other gas environment)